This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
High speed and precision, non-contact dispensing of fluids to multi-well plates.
- Partner:The University of Manchester
- Facility:Manchester Institute for Biotechnology
Or call us now on 0161 275 8382
Detailed Description
High precision, non-contact, high speed, low volume (2.5nl) droplet transfer of diverse fluids from 384 and 1536-well microplates to destination plates or slides supporting a range of applications in drug discovery and basic research.
Uses/Applications
A broad range of applications including: miniaturization of sample and reagent transfer for high throughput assays and DNA library generation; Cell based and biochemical assays; micro-sampling; protein crystallography; gene expression and genotyping; qPCR; sequencing and functional screening assays.
Drop volume: 2.5 nL
Volume transfer range: 2.5 nL – 5 µL (dependent on fluid type, assay, and source plate used)
Reagents supported: DMSO, PCR and qPCR reagents, synthetic biology reagents, NGS reagents, primers and probes, cell culture media, proteins, nucleic acids, up to 50% glycerol, serum, plasma, antibodies, enzymes, and protein crystallography reagents
Transfer accuracy: <10% deviation from target volume
Transfer precision: <8% CV (except protein crystallography fluid class)
Source labware compatibility: All Echo Qualified 384- and 1536-well Microplates
Destination labware compatibility: All Echo Qualified Microplates and most ANSI-compliant/SBS-standard microplates in 96-, 384-, 1536-, and 3456-well formats 8 to 16 mm well height